Port-a-Patch
Patchliner
SynchroPatch 96
Lateral Flow Products
BioSensor
Biochip Products
Solid Transfer
Products
LM10 / LM20
NS500
Qsep100
SensiQ
SensiQ Pionner
SensiQ Discovery
Monolith NT. 115
Discovery 2500
system
Discovery
Adcon Reader
Analytical SFC
Preparative SFC
Extraction
& Particle Designs
Magnetic Assay
Reader
Nanion Technologies»çÀÇ Àü±â»ý¸®ÇÐÀû ½ÇÇèÀ» À§ÇÑ ÀÌ ÀåÄ¡´Â 1°³ÀÇ ¼¼Æ÷¸¦ ÀÚµ¿À¸·Î ÈíÂøÇÏ¿© Ưº°ÇÑ ¼¼Æ÷ Á¶ÀÛ
±â¼ú ¾øÀÌ Ion Channel ¿¬±¸¸¦ ÇÒ ¼ö ÀÖ½À´Ï´Ù.
- easy to use complete patch clamp setup
- glass chip for true giga-ohm seals
- accessible for fast fluid exchange
Conventional Patch Clamp Nanionȍ˂ Port-a-Patch
composed of many pieces one entity device
Trained personal necessary Technician operated
Anti-vibration table necessary Inherently stable
Large liquid consumption : > 50ml Small liquid consumption : <10ul
Elaborate fast fluid exchange Effortless exchange of internal solution
Low throughput : < 15 data point per day Higher throughput : up to 50 data point per day
Specification
- Chip resistance : 2~5M§Ù
- Seal resistance : > 1G§Ù
- Whole cell resistance : > 200M§Ù
- Series resistance : < 15M§Ù
- Liquid consumption : 10§¡ / compound
- Perfusion time constant : < 10ms
- Avg. whole cell stability : ~ 20min
- Successful whole cell recording : 30~50 % (CHO/HEK)
- Throughput : 50 data point /day